Bringing Specimens into Focus with the Compound Light Microscope
by Professor Stephen T. Abedon Steve Abedon, Ph.D., has taught well over 100 individual classes and ~10 individual courses (including introductory biology, microbiology, physiology, biochemistry, etc.) over his 30+ years as a college professor. He has over 150 publications, including multiple monographs, and is ranked in the top 1 or 2 percent among researchers of all disciplines worldwide. His research focuses on bacteriophage biology and phage therapy. (abedon.1@osu.edu)
phage.org | phage-therapy.org | biologyaspoetry.org | abedon.phage.org | google scholar | publications
Jump to: 🔬 The Microscope | 🎛️ Using It | 🎥 Focusing In | 📏 Calculators | 🔢 Petroff-Hausser | 🔍 Types of Microscopy | 🧮 More Calculators
biologyaspoetry.org/laboratories/microscopy/ · Abedon’s Books
How can I improve this page? contact: microscopy@biologyaspoetry.org
A compound microscope magnifies in two stages. An objective lens, close to the specimen, makes an enlarged image, and an ocular lens, at your eye, enlarges that image again. What you see is also only as good as the light that reaches the specimen, so the parts of the microscope are easiest to understand in the order that light passes through them. Start with the video, in which I walk through a typical teaching microscope, and then use the rest of this tab as a reference.
The parts below are listed in the order in which light passes through them, from the lamp in the base up to your eyes.
Two further controls do not sit in the light path but move the specimen within it: the coarse focus knob, for large movements of the stage, and the fine focus knob, for small ones. On these microscopes they move the stage up and down rather than moving the objectives.
Total magnification is the objective’s magnification multiplied by the ocular’s. With 10× oculars:
| Objective | Total magnification | Also called | Used for |
|---|---|---|---|
| 4× | 40× | scanning objective | finding the specimen and centering it; always start here |
| 10× | 100× | low power | tissues, larger protists, getting oriented |
| 40× | 400× | high power, “high-dry” | individual cells and their larger parts; the usual stopping point for introductory work |
| 100× | 1000× | oil immersion | bacteria and fine detail; requires a drop of immersion oil |
Magnification by itself is not the goal. What matters is how much detail can be told apart, that is, resolution, and past a certain point additional magnification makes the image larger without making it any sharper. The Calculators tab works out where that point lies for each objective.
Your answers are kept as you go. In most browsers they are still here if you leave and come back on the same device, though a private window, a different device, or cleared browser data will start you fresh. When you are done, download a summary to hand in or keep. It lists your answers and, for each, whether it was right the first time.
Most of the trouble that people have with microscopes comes from a small number of settings, and nearly all of it can be avoided by starting from the same place every time. Below are the starting settings, the sequence for bringing a specimen into focus, and then what to check when things go wrong, organized by what you are (or are not) seeing.
Once the 10× or 40× objective is in place, there should be no need for the coarse focus knob at all. At those magnifications it moves the stage too far too quickly, and at 40× and above it can drive the slide into the objective.
The iris diaphragm, not the lamp, is the control to reach for first. Unstained specimens such as cheek cells and most living microorganisms are nearly transparent, and with the iris diaphragm wide open they simply disappear into the glare. Closing it makes them visible. The amount of light you want coming up through the condenser is usually just a little more than almost none at all.
If the light is too intense for your eyes, try closing the iris diaphragm further before reducing the intensity at the illuminator, though feel free to do that too. There is a cost to closing the iris diaphragm very far, however: past a point, fine detail blurs even as contrast improves. The Calculators tab shows why.
If the condenser belongs all the way up, why can it be lowered at all? Because the best height depends on the objective and on the specimen, and experienced microscopists do adjust it. When starting out, though, all the way up is the place to put it and to leave it. A condenser lowered at 4× and then forgotten is likely to cause trouble later, since the 40× and 100× objectives are the ones that most need everything the condenser can deliver, both light and resolution. Its height, in other words, is a setting worth understanding before it is a setting worth changing. The usual reasons for lowering it are these:
Whatever the reason for lowering the condenser, return it to the top before moving on to 40× or 100×.
When I come to a microscope where someone is having trouble, I check the same things in roughly the same order: whether there is enough light coming from the lamp, whether the objective is anywhere near the slide, whether what is in view is actually on the slide, and whether too much light is coming up through the condenser. Find what you are seeing below.
This is the laboratory itself. In each of these short videos I bring a specimen into focus the way that the Using It tab describes, narrating what I am doing as I do it, up to high magnification. Watch one, then do the same thing yourself, ideally with the same kind of specimen.
There are 31 of them, grouped roughly from easiest to hardest: stained, prepared slides first, then unstained wet mounts, then protists and algae, and finally bacteria, which call for oil immersion. Click a thumbnail to play its video here on the page; it opens full width, and opening another closes the first.
Jump to: prepared slides | wet mounts | protists and algae | bacteria
Commercially prepared, thin, and stained, these are the easiest specimens to find and focus, so they are the place to start. Staining supplies the contrast, so more light is acceptable here than with the specimens that follow.
Living or freshly prepared material in a drop of water under a coverslip. Most of these are unstained and nearly transparent, so the iris diaphragm matters more here than anywhere else on this page, and the thinner the specimen, the better it will look.
Single-celled and colonial eukaryotes, some on prepared slides and some living. Several are large enough to find easily at 4×, and several reward going to 400× for internal detail.
The smallest specimens on this page, at roughly a micrometer across. Individual bacteria are resolved well only with the oil-immersion objective, which makes these the hardest to bring into focus, and the most satisfying.
These are not mine, but they are worth the time. The first two cover making wet mounts and using a microscope; the rest are simply the microscopic world, beautifully filmed.
Four questions that come up at the microscope: how much is this magnified, how wide is the circle I am looking at, how big is the thing in it, and how fine a detail can this objective show? The first three depend only on the lenses. The fourth depends on the light and on the iris diaphragm, and it explains why about 1000× is as far as it is worth taking a light microscope.
The defaults describe a typical teaching microscope. If your oculars or objectives differ, change them here; everything else on this tab updates. An objective’s numerical aperture (NA) is printed on its barrel after the magnification, as in “40/0.65.” An ocular’s field number is printed after its magnification, as in “WF10×/18.”
| Objective | Magnification (×) | Numerical aperture | Immersion |
|---|---|---|---|
| Scanning | air | ||
| Low power | air | ||
| High-dry | air | ||
| Oil immersion | oil |
To measure the field yourself, place a clear millimeter ruler on the stage, focus on its marks with the lowest-power objective, and count how many millimeters span the circle. Fields at higher magnifications are then smaller in proportion. (Measuring directly at high power this way does not work well; the ruler’s marks are too far apart.)
Estimate the size of something in view by comparing it with the width of the field. Judge how many of it, laid end to end, would span the circle from edge to edge. Decimals are fine: if one cell spans a quarter of the field, enter 4; if it spans two-thirds of it, enter 1.5.
For comparison, roughly: a typical rod-shaped bacterium such as Escherichia coli, 1 to 3 µm long; a human red blood cell, 7 to 8 µm across; a human cheek cell, 50 to 60 µm; a Paramecium caudatum, 200 to 300 µm long.
Total magnification is objective × ocular. Field of view is the ocular’s field number divided by the objective’s magnification. It does not depend on the ocular’s magnification, because the field number already describes the ocular. If you measured the field instead, the field at any other objective is your measurement scaled by the ratio of the two objectives’ magnifications.
Resolution is the smallest distance at which two points can still be told apart as two. Following Ernst Abbe, who worked out the theory in the 1870s, it is estimated here as
d = λ / (NAobjective + NAcondenser)
where λ is the wavelength of the light and NA is numerical aperture, a measure of how wide a cone of light a lens gathers. With the condenser matched to the objective this becomes the familiar d = λ/2NA. A related and commonly quoted criterion, due to Lord Rayleigh, gives d = 0.61λ/NA, which yields similar values. Both are idealized: real specimens, real lenses, and real alignment generally do somewhat worse.
Blue filters. Because d is proportional to λ, shorter wavelengths resolve finer detail, and a filter that passes mainly blue light, around 450 to 490 nm, improves resolution by roughly 10 to 20% compared with green; choose “blue filter” above to see the difference. Not every blue filter does this, however. The blue filter found beneath the condenser of many teaching microscopes is often a “daylight” filter, meant mainly to correct the yellowish color of a halogen lamp, and it still passes a broad range of colors, so its effect on resolution is expected to be modest. Blue light also looks dimmer, because the eye is less sensitive to it.
Numerical aperture is also why oil immersion works. A lens in air cannot have an NA above 1.0, and in practice dry objectives stop near 0.95. Immersion oil has nearly the same refractive index as glass, so light that would otherwise bend away at the glass–air boundary is instead collected by the objective, allowing NAs of 1.25 to 1.4.
Useful magnification is the range, by a widely used rule of thumb about 500 to 1000 times the objective’s NA, over which magnification enlarges resolved detail enough for your eye to see it comfortably. Below that range the lenses resolve more than you can see; above it, the image gets bigger without getting sharper, what is called empty magnification. With the best oil-immersion objectives, at NA 1.25 to 1.4, the range tops out between about 1250× and 1400×, and that is the reason why light microscopes rarely go past 1000× to 1500×.
The iris diaphragm trade-off. Closing the iris diaphragm narrows the cone of light coming from the condenser, which lowers NAcondenser. Contrast and depth of field improve, which is why closing it makes transparent cells visible, but d grows, so the finest detail blurs. Try the slider above. The best setting is a compromise, and it differs from specimen to specimen, which is why it is worth adjusting rather than leaving where you found it.
A counting chamber turns a microscope into a measuring instrument. It is a thick glass slide with a grid etched into a central platform. The platform sits a precise distance below the coverslip, so every square of the grid holds a known volume of liquid. Count the cells over some number of squares, divide by the volume, and you have the number of cells per milliliter.
The Petroff-Hausser chamber is made for bacteria. Its platform is only 0.02 mm below the coverslip, shallow enough that small cells stay close to a single focal plane. The more familiar hemocytometer uses essentially the same grid at a depth of 0.1 mm, which suits larger cells such as yeast and blood cells. The calculator below handles either.
The ruled area of a Petroff-Hausser chamber is a square 1 mm on a side. It is divided into 25 groups of squares, 5 by 5, and each group is divided again into 16 small squares, 4 by 4, for 400 small squares in all.
| Counting unit | Side | Area | Volume at 0.02 mm deep | Cells/mL = |
|---|---|---|---|---|
| small square | 0.05 mm | 1/400 mm² | 5 × 10−8 mL | mean count × 2 × 107 |
| 10 small squares | – | 1/40 mm² | 5 × 10−7 mL | count × 2 × 106 |
| group of 16 | 0.2 mm | 1/25 mm² | 8 × 10−7 mL | mean count × 1.25 × 106 |
| whole ruled area | 1 mm | 1 mm² | 2 × 10−5 mL | count × 5 × 104 |
Each result is then multiplied by any dilution made before loading. The conversions rest on 1 mm³ = 1 µL = 10−3 mL.
Cells on the lines. A cell touching a boundary could be counted in either of two squares, or in neither. A common convention is to count cells that touch the top or left boundary of a square but not those touching the bottom or right boundary. Whatever convention you use, use it consistently, so that no cell is counted twice. The practice field at the bottom of this tab uses this one.
Why a few hundred? Cells land in squares at random, so counts vary from square to square even in a perfectly mixed sample. For randomly scattered cells, the uncertainty of a total count N is roughly √N, a relative uncertainty of about 1/√N: 20% for 25 cells, 10% for 100, and 5% for 400.
Total versus viable. A counting chamber counts every cell it can see, alive or dead. A plate count, by contrast, counts only cells able to form colonies. The two therefore answer different questions, and a chamber count is expected to be at least as high as a plate count of the same sample.
Counting 10 small squares? Either enter the 10 counts, one per square, with “small square” chosen above, or enter their total as a single count, with “several small squares (total)” chosen and set to 10. Entering the counts square by square also lets the calculator check whether they vary more than chance alone would explain.
Every kind of microscopy has to solve two problems: making small things big enough to see (magnification and resolution) and making them stand out from their surroundings (contrast). Brightfield microscopy, the kind used on the rest of this page, is simple and inexpensive, but it gets its contrast mostly from staining, and staining generally kills cells. The other methods below differ mainly in how they create contrast, what they can resolve, and what they require of the specimen. Resolutions given are rough, typical values.
Light passes through the specimen, which is seen as darker or colored against a bright background.
Two separate optical paths, one per eye, give a three-dimensional view of the surface of an object, usually lit from above.
A special condenser lights the specimen only from oblique angles, so only light scattered by the specimen enters the objective. Specimens glow against black.
Converts the tiny delays that light suffers passing through denser parts of a transparent specimen into differences in brightness. Invented by Frits Zernike, for which he received the 1953 Nobel Prize in Physics.
Also called Nomarski microscopy, after Georges Nomarski. Polarized light is split into two slightly offset beams that are recombined after passing through the specimen, giving an image that looks shaded, almost three-dimensional.
Specimens are labeled with molecules that absorb light of one color and emit another, and filters pass only the emitted light. Labels include fluorescent dyes, antibodies carrying dyes, and fluorescent proteins such as GFP.
A fluorescence method in which a pinhole blocks light from outside the plane of focus. Scanning the specimen point by point builds up sharp optical sections, which can be stacked into three dimensions. The principle was patented by Marvin Minsky, who filed in 1957.
A family of fluorescence methods (among them STED, PALM, STORM, and structured illumination) that get around the diffraction limit described on the Calculators tab by controlling when and where labels fluoresce. Eric Betzig, Stefan Hell, and William Moerner shared the 2014 Nobel Prize in Chemistry for this work.
Electrons have far shorter wavelengths than visible light, so electron microscopes can resolve far smaller things. The first was built by Ernst Ruska and Max Knoll in the early 1930s; Ruska received a share of the 1986 Nobel Prize in Physics. The price is that specimens must be in a vacuum, which rules out living cells.
Electrons pass through a very thin section or a thin layer of particles. Contrast comes from heavy-metal stains, including negative staining, in which the stain surrounds rather than enters small particles.
A focused beam of electrons scans across the surface of a specimen, usually coated with a thin layer of metal, and electrons given off from the surface build the image.
Specimens are frozen so quickly that water becomes glass rather than ice, and are imaged unstained while frozen. Thousands of images of identical particles can be combined into three-dimensional structures. Jacques Dubochet, Joachim Frank, and Richard Henderson shared the 2017 Nobel Prize in Chemistry for its development.
No lenses or beams at all: an extremely fine tip on a flexible arm is dragged or tapped across a surface, and its deflections map the surface’s height.
Robert Hooke described and named cells, in cork, in his Micrographia of 1665, using a compound microscope. Antonie van Leeuwenhoek, using single-lens microscopes of his own making, soon after became the first to report protists and bacteria. For much of the next two centuries, however, compound microscopes were limited more by the imperfections of their lenses than by the physics of light, and it was not until Abbe’s work in the 1870s that the ultimate limit of light microscopy, described on the Calculators tab, was understood.