🎓 Doing Microscopy

Bringing Specimens into Focus with the Compound Light Microscope

by Professor Stephen T. Abedon Steve Abedon, Ph.D., has taught well over 100 individual classes and ~10 individual courses (including introductory biology, microbiology, physiology, biochemistry, etc.) over his 30+ years as a college professor. He has over 150 publications, including multiple monographs, and is ranked in the top 1 or 2 percent among researchers of all disciplines worldwide. His research focuses on bacteriophage biology and phage therapy. (abedon.1@osu.edu)

phage.org | phage-therapy.org | biologyaspoetry.org | abedon.phage.org | google scholar | publications

Jump to:   🔬 The Microscope  |  🎛️ Using It  |  🎥 Focusing In  |  📏 Calculators  |  🔢 Petroff-Hausser  |  🔍 Types of Microscopy  |  🧮 More Calculators

What Is Doing Microscopy? The microscope is perhaps the instrument most identified with biology, and it is also easy to misuse. Nearly everything that goes wrong at the bench comes down to a few settings: too much light, a stage nowhere near the objective, a slide resting on top of its clip rather than in it. This page is about getting those right, with the emphasis on the compound brightfield microscope of the introductory biology laboratory. The Microscope follows the light from lamp to eye. Using It gives the starting settings, the sequence for bringing a specimen into focus, and what to check when a specimen will not cooperate. Focusing In, the laboratory proper, is a set of 31 short videos in which I bring prepared slides, wet mounts, protists, and bacteria into focus at high magnification.

Calculators works out total magnification, the width of the field of view, the sizes of the things in it, and why around 1000× is about as far as a light microscope usefully goes. Petroff-Hausser turns the microscope into a way of counting cells, with a calculator and a practice field. Types of Microscopy comes last: brightfield is only one of many ways of seeing the very small, and each of the others trades something for what it gains.
Your Practice Counting answers are kept in your browser as you go, and The Microscope tab has a button to download a summary of your work. To share a particular section, use the 🔗 link beside its heading, which copies a web address that opens the page right there. To cite this tool: Abedon, S.T. (2026). Doing Microscopy. biologyaspoetry.org/laboratories/microscopy/.

biologyaspoetry.org/laboratories/microscopy/  ·  Abedon’s Books

How can I improve this page?  contact: microscopy@biologyaspoetry.org

🔬 The Compound Microscope

A compound microscope magnifies in two stages. An objective lens, close to the specimen, makes an enlarged image, and an ocular lens, at your eye, enlarges that image again. What you see is also only as good as the light that reaches the specimen, so the parts of the microscope are easiest to understand in the order that light passes through them. Start with the video, in which I walk through a typical teaching microscope, and then use the rest of this tab as a reference.

How to Use a Microscope
From plugging in the microscope to putting it away: the light path, the ocular adjustments, the four objective lenses, finding and focusing a specimen, oil immersion, and what I check first when a student cannot get a specimen into focus. About 14 minutes.

💡 Following the Light

The parts below are listed in the order in which light passes through them, from the lamp in the base up to your eyes.

  1. Illuminator. The light source, in the base, with an on/off switch and an intensity control. Start with the intensity at or near its maximum.
  2. Condenser. A lens beneath the stage that gathers light and concentrates it onto the specimen. It should be all the way up. (For why it can be lowered at all, see Why the Condenser Moves.)
  3. Iris diaphragm. An adjustable opening in the condenser, set with a small lever. It is the main control over how much light reaches the specimen and, just as important, over contrast. For most specimens it should be almost, but not completely, closed.
  4. Stage. The platform on which the slide rests. A mechanical stage holds the slide in a spring clip and moves it with two knobs, one for each direction.
  5. Objective lenses. On modern microscopes, usually four, on a revolving nosepiece: 4×, 10×, 40× (the “high-dry” objective), and 100× (the oil-immersion objective).
  6. Ocular lenses. The two eyepieces, each 10×. The distance between them adjusts to the distance between your eyes, and one of them can be focused separately to allow for a difference between your two eyes.

Two further controls do not sit in the light path but move the specimen within it: the coarse focus knob, for large movements of the stage, and the fine focus knob, for small ones. On these microscopes they move the stage up and down rather than moving the objectives.

Labeled diagram of a compound light microscope
The parts of a compound light microscope.

🔍 Magnification

Total magnification is the objective’s magnification multiplied by the ocular’s. With 10× oculars:

ObjectiveTotal magnificationAlso calledUsed for
4×40×scanning objectivefinding the specimen and centering it; always start here
10×100×low powertissues, larger protists, getting oriented
40×400×high power, “high-dry”individual cells and their larger parts; the usual stopping point for introductory work
100×1000×oil immersionbacteria and fine detail; requires a drop of immersion oil

Magnification by itself is not the goal. What matters is how much detail can be told apart, that is, resolution, and past a certain point additional magnification makes the image larger without making it any sharper. The Calculators tab works out where that point lies for each objective.

📖 A Few Terms

🧭 What Is on Each Tab

💾 Your Work

Your answers are kept as you go. In most browsers they are still here if you leave and come back on the same device, though a private window, a different device, or cleared browser data will start you fresh. When you are done, download a summary to hand in or keep. It lists your answers and, for each, whether it was right the first time.

    🎛️ Using It

    Most of the trouble that people have with microscopes comes from a small number of settings, and nearly all of it can be avoided by starting from the same place every time. Below are the starting settings, the sequence for bringing a specimen into focus, and then what to check when things go wrong, organized by what you are (or are not) seeing.

    ✅ Before You Look

    • Make sure the microscope is plugged in and turned on. Start with the light at its highest intensity; you can step it down later if it is too bright for your eyes.
    • Make sure the condenser is all the way up, and leave it there. (There are reasons to lower it, described below, but none that should concern you at first.)
    • Make sure the iris diaphragm is almost completely closed, but not completely.
    • Rotate the 4× objective into place, and raise the stage (and thereby the focus) to its highest position.
    • Turn the adjustable ocular (on these microscopes, the right one) all the way down, so that it starts out matching the other one.
    • Adjust the distance between the oculars, the interpupillary distance, until you see a single circle comfortably with both eyes. Most modern microscopes are binocular, and using both eyes is the proper way to use them.
    • Make sure that your slide is actually in the clip. Not on top of the clip, not beneath the clip, not at some crazy angle relative to the clip, but in the clip, resting flat on the stage with the clip holding it in place. The coverslip, if there is one, faces up.

    🎯 Bringing a Specimen into Focus

    1. Using the mechanical stage knobs, move the slide so that the specimen sits directly under the 4× objective. Move slides with the knobs, not by hand (unless of course you do not have a mechanical stage).
    2. Looking through the oculars, lower the stage slowly with the coarse focus knob until something comes into sharp focus. It will not take much. Starting with the stage at its highest position means that you will usually be moving the specimen away from the objective as you focus, which is the safe direction.
    3. Move the slide a little with the stage knobs. If what you are looking at moves, it is on the slide. If it does not move, you have focused on something else, such as dust on a lens, and should start over.
    4. Center whatever you want to look at in the field of view.
    5. Switch to the 10× objective. Because the objectives are parfocal, the specimen should be nearly in focus; sharpen it with the fine focus knob only.
    6. Center it again, then switch to the 40× objective and again use only the fine focus. You may need to adjust the focus a bit more here than before, since depth of field shrinks as magnification increases and optics are not perfect. What was cleanly in focus will not necessarily remain in focus, but it should be close.
    7. If the specimen calls for it, go on to oil immersion, below. Otherwise, 400× is usually where to stop.

    Once the 10× or 40× objective is in place, there should be no need for the coarse focus knob at all. At those magnifications it moves the stage too far too quickly, and at 40× and above it can drive the slide into the objective.

    💧 Oil Immersion (1000×)

    1. Get the specimen focused and centered with the 40× objective first.
    2. Rotate the nosepiece partway, so that the 40× objective has moved aside but the 100× objective is not yet in place.
    3. Place one drop of immersion oil on the slide, right where the light comes through.
    4. Rotate the 100× objective into the oil. Do not rotate back through the oil with the 40× objective, which is designed to be used dry.
    5. Sharpen the focus with the fine focus knob. You will often need more light at this magnification; open the iris diaphragm a little.
    6. When you are done, wipe the oil from the 100× objective with lens paper.
    Never let immersion oil touch any other objective. Only the 100× objective is made to be used in oil. Oil on the front of a dry objective, most often the 40×, smears its image until it is cleaned off, and oil left there can work its way into the lens. Once there is oil on a slide, do not swing the 40× objective back over it; if you need a lower magnification, rotate the nosepiece the other way, to the 4× or 10× objective, whose longer working distances keep them clear of the oil.

    🔅 Light, and Why Less Is Often More

    The iris diaphragm, not the lamp, is the control to reach for first. Unstained specimens such as cheek cells and most living microorganisms are nearly transparent, and with the iris diaphragm wide open they simply disappear into the glare. Closing it makes them visible. The amount of light you want coming up through the condenser is usually just a little more than almost none at all.

    If the light is too intense for your eyes, try closing the iris diaphragm further before reducing the intensity at the illuminator, though feel free to do that too. There is a cost to closing the iris diaphragm very far, however: past a point, fine detail blurs even as contrast improves. The Calculators tab shows why.

    ↕️ Why the Condenser Moves

    If the condenser belongs all the way up, why can it be lowered at all? Because the best height depends on the objective and on the specimen, and experienced microscopists do adjust it. When starting out, though, all the way up is the place to put it and to leave it. A condenser lowered at 4× and then forgotten is likely to cause trouble later, since the 40× and 100× objectives are the ones that most need everything the condenser can deliver, both light and resolution. Its height, in other words, is a setting worth understanding before it is a setting worth changing. The usual reasons for lowering it are these:

    Whatever the reason for lowering the condenser, return it to the top before moving on to 40× or 100×.

    ⚠️ When Things Go Wrong

    When I come to a microscope where someone is having trouble, I check the same things in roughly the same order: whether there is enough light coming from the lamp, whether the objective is anywhere near the slide, whether what is in view is actually on the slide, and whether too much light is coming up through the condenser. Find what you are seeing below.

    I see nothing, or only a bright blur
    • Is the lamp on and turned up? If almost no light is coming from the illuminator, turn it up.
    • Is the specimen actually under the objective? Go back to 4× and center it using the stage knobs, watching from the side if need be.
    • Try the finger test. At 10× or 40×, if you can fit a finger between the objective and the slide, whatever you are looking at is not in focus; the stage is far too low.
    • Is the iris diaphragm wide open? Too much light can make an unstained specimen invisible. Close it most of the way.
    Something is sharp, but it does not move when I move the slide
    • You have focused on something other than the specimen, often dust or a smudge on an ocular or objective, or the surface of the coverslip. Start over: stage all the way up at 4×, then lower it slowly.
    • If it is still there after you have cleaned the oculars and the objective with lens paper, and it still does not move with the slide, ask for help rather than taking anything apart.
    Everything looks washed out, and unstained cells are invisible
    • Close the iris diaphragm. This is the single most common fix.
    • Close the iris diaphragm rather than turning down the lamp. Dimming the lamp darkens the image without adding contrast.
    At 4× there is a bright spot in the middle, or the edges are dark
    • The condenser, all the way up, may be concentrating the light into a smaller area than the wide 4× field. This is harmless for finding and centering a specimen, which is all 4× is for here.
    • If it bothers you, lower the condenser slightly, or swing out its top lens if it has one, but return it to the top before switching to a higher-magnification objective. See Why the Condenser Moves.
    It was in focus, and then I switched objectives and lost it
    • Was the specimen centered first? The field of view at 40× is a tenth the diameter of the field at 4×, so anything off-center disappears.
    • Use only the fine focus after switching. If the specimen still will not come back, return to the previous objective, recenter, refocus, and switch again.
    I can focus at 4× and 10×, but never at 40×
    • Check that the slide is not upside down. With the coverslip facing down, the slide itself is thicker than the working distance of the 40× objective, and the specimen cannot be reached.
    • Check that the slide is in the clip rather than resting on top of it, which lifts one end.
    • A thick wet mount, or a coverslip with too much liquid under it, can have the same effect. Thinner is better.
    At 1000× the image is dim or fuzzy
    • Is there oil between the 100× objective and the slide? Without it, the objective cannot give a sharp image.
    • An air bubble in the oil can have the same effect; rotate the objective out and back in.
    • Open the iris diaphragm a little; this magnification usually needs more light.
    • Check that the condenser is all the way up. A lowered condenser costs the 100× objective both light and resolution.
    • If the 40× objective was rotated through the oil, its image will be smeared until it is cleaned.
    I see two overlapping circles, or my eyes tire quickly
    • Readjust the interpupillary distance until the two circles merge into one.
    • Refocus the adjustable ocular: with your eye on the fixed ocular, sharpen the image with the fine focus; then, with your eye on the adjustable one, sharpen it by turning that ocular.
    • Keep both eyes open and relaxed, as though looking at something far away.
    Everything looks hazy no matter what I do
    • Dirty optics are not pleasant to look through. Clean the oculars and objectives with lens paper and lens-cleaning fluid, never with paper towels or tissues, which can scratch.
    • Check whether oil has found its way onto a dry objective or onto the top of the condenser.

    📦 Putting It Away

    • Remove the slide, and clean up as best you can, including any oil. Use lens paper on the lenses.
    • Turn the microscope off at its switch, then unplug it and wrap the cord.
    • Carry the microscope with two hands, one on the arm and the other under the base. Never pick it up by the upper part, which holds the oculars; on these microscopes it can pop right off.

    🎥 Focusing In

    This is the laboratory itself. In each of these short videos I bring a specimen into focus the way that the Using It tab describes, narrating what I am doing as I do it, up to high magnification. Watch one, then do the same thing yourself, ideally with the same kind of specimen.

    There are 31 of them, grouped roughly from easiest to hardest: stained, prepared slides first, then unstained wet mounts, then protists and algae, and finally bacteria, which call for oil immersion. Click a thumbnail to play its video here on the page; it opens full width, and opening another closes the first.

    Jump to: prepared slides  |  wet mounts  |  protists and algae  |  bacteria

    Watching for technique. Pay attention to what happens between the moments when the specimen is in focus: how far the stage moves at 4× compared with 40×, recentering before switching objectives, and adjustments to the light. Those are the habits that make the difference at your own microscope.

    🧫 Prepared Slides: Tissues and a Small Animal

    Commercially prepared, thin, and stained, these are the easiest specimens to find and focus, so they are the place to start. Staining supplies the contrast, so more light is acceptable here than with the specimens that follow.

    Human pancreas
    The pancreas is both a digestive and an endocrine organ. Look for the islets of Langerhans, clusters of hormone-producing cells that typically stain paler than the surrounding digestive-enzyme-producing tissue.
    Thyroid and parathyroid
    Thyroid tissue consists of round follicles, rings of cells surrounding a stored secretion (colloid); parathyroid tissue by contrast is a dense mass of cells.
    Note that at a little after 2:30 I say “atomic nuclei” when I should have said “cell nuclei.” Oops…
    Human adrenal gland
    An outer cortex and an inner medulla, which make different hormones and differ in appearance.
    Thymus, human fetus (developing thymus)
    The thymus is where T cells mature. Look for lobules, each with a darker outer region and a paler center.
    Pig skin
    A layered epidermis at the surface over a thicker dermis. (Problematic ending.)
    Hydra with bud
    A small freshwater animal related to jellyfish, here reproducing asexually by budding: the bud is a new individual growing from the side of its parent.

    💧 Wet Mounts: Cells You Can Prepare Yourself

    Living or freshly prepared material in a drop of water under a coverslip. Most of these are unstained and nearly transparent, so the iris diaphragm matters more here than anywhere else on this page, and the thinner the specimen, the better it will look.

    My cheek cells
    Flat epithelial cells scraped from the inside of the cheek, each with a single nucleus. Isolated cells tend to be easier to see than clumps.
    Elodea
    A pond plant with leaves only a few cells thick. Look for cell walls and many green chloroplasts, which in living cells can often be seen moving as the cytoplasm streams.
    Red pepper
    The red color is carried in chromoplasts, pigment-containing plastids related to chloroplasts.
    Potato
    Potato cells are packed with starch grains, which under good conditions show faint concentric layering.
    Purple onion
    Epidermal cells whose large central vacuoles are colored by water-soluble pigments, making the vacuole itself visible.

    🌿 Protists and Algae

    Single-celled and colonial eukaryotes, some on prepared slides and some living. Several are large enough to find easily at 4×, and several reward going to 400× for internal detail.

    Euglena (prepared slide)
    A single-celled, flagellated protist with chloroplasts. Look also for its red eyespot.
    Euglena, a eukaryotic alga
    A second look at Euglena.
    Volvox, a eukaryotic alga
    Hollow spherical colonies of hundreds to thousands of flagellated cells, often with daughter colonies developing inside.
    Volvox zygote
    The thick-walled products of Volvox sexual reproduction.
    Spirogyra conjugation
    A filamentous green alga named for its spiral, ribbon-shaped chloroplasts. For simplicity, focus especially on Spirogyra before conjugation.
    Diatoms
    Algae enclosed in intricately patterned walls made of silica, the main ingredient of glass.
    Ceratium, a dinoflagellate
    A eukaryotic alga armored with plates and extended into long horns.
    Paramecium caudatum
    A slipper-shaped ciliate, a single cell covered in cilia, with an oral groove along one side.
    Stentor (prepared slide)
    A very large, trumpet-shaped ciliate; look for its long macronucleus, which resembles a string of beads.
    Stentor, a ciliate
    A second look at Stentor.
    Spirostomum
    A long, worm-like ciliate that can contract remarkably quickly.
    Trypanosoma gambiense
    A flagellated parasite, the cause of African sleeping sickness (now usually named Trypanosoma brucei gambiense), usually seen in blood smears among the far more numerous red blood cells of its host.
    Radiolaria
    Protozoa that live within intricate skeletons, called tests, made mostly of silica.

    🦠 Bacteria, Including Cyanobacteria

    The smallest specimens on this page, at roughly a micrometer across. Individual bacteria are resolved well only with the oil-immersion objective, which makes these the hardest to bring into focus, and the most satisfying.

    Typical cocci
    Spherical bacteria. Note how they are arranged: singly, in pairs, in chains, or in clusters.
    Spirillum
    Rigid, spiral-shaped bacteria.
    Typical bacillus (Bacillus megaterium)
    Rod-shaped bacteria; B. megaterium is among the larger ones, which makes it a good first bacillus.
    Bacillus anthracis-infected tissue
    The cause of anthrax, seen as rods, often in chains, among the host’s own cells.
    Nostoc, a cyanobacterium
    Bead-like chains of cells. Look for the occasional larger, paler cells, called heterocysts, which fix nitrogen.
    Oscillatoria, a cyanobacterium
    Long filaments of disk-shaped cells, without heterocysts. Living filaments glide and sway, which is the source of the name.
    From elsewhere: other microscopy videos worth watching

    These are not mine, but they are worth the time. The first two cover making wet mounts and using a microscope; the rest are simply the microscopic world, beautifully filmed.

    Making Wet Mounts
    How to make wet mounts like those in the Focusing In videos. Note the consistent emphasis on making sure that what is being viewed is very thin!
    BIOLOGY 10 – Basic Microscope Setup and Use
    A short introduction to using a compound microscope. A few details will differ between different microscopes; for instance, the highest-magnification objective shown may be 40× rather than 100×.
    Pond Scum Under the Microscope – Pondlife, Episode #1
    A nice place to start when considering the microscopic, as it is fun, historical, informative, and pretty.
    The Microscopic Life in a Pond – 1000× Magnification
    A very pretty and reasonably informative exploration of pond water. Beware various typos, though.
    Life in a Drop of Water
    Gets a bit slow at times and isn’t always terribly informative, but still it’s pretty wonderful to watch.
    Real-Time Movie of Cell Locomotion of Amoeba proteus
    An amoeba, indeed Amoeba proteus, moving.
    How Microscopic Hunters Get Their Lunch
    Various heterotrophic eukaryotic microorganisms obtaining nutrients, particularly their carbon and energy sources.
    Stentors: Single-Celled Giants
    Nearly the entire video is on Stentor.
    Lacrymaria: Vicious Long-Necked Predators
    On the ciliate Lacrymaria.

    📏 Calculators

    Four questions that come up at the microscope: how much is this magnified, how wide is the circle I am looking at, how big is the thing in it, and how fine a detail can this objective show? The first three depend only on the lenses. The fourth depends on the light and on the iris diaphragm, and it explains why about 1000× is as far as it is worth taking a light microscope.

    🔬 Your Microscope

    The defaults describe a typical teaching microscope. If your oculars or objectives differ, change them here; everything else on this tab updates. An objective’s numerical aperture (NA) is printed on its barrel after the magnification, as in “40/0.65.” An ocular’s field number is printed after its magnification, as in “WF10×/18.”

    ObjectiveMagnification (×)Numerical apertureImmersion
    Scanningair
    Low powerair
    High-dryair
    Oil immersionoil

    To measure the field yourself, place a clear millimeter ruler on the stage, focus on its marks with the lowest-power objective, and count how many millimeters span the circle. Fields at higher magnifications are then smaller in proportion. (Measuring directly at high power this way does not work well; the ruler’s marks are too far apart.)

    📐 Magnification, Field of View, and Resolution

    For white light, 550 nm, green, is the usual choice: it lies near the middle of the visible range, where the eye is most sensitive. See below on blue filters.
    Now: 1.00. Closing the iris diaphragm moves this toward 0.

    📏 How Big Is It?

    Estimate the size of something in view by comparing it with the width of the field. Judge how many of it, laid end to end, would span the circle from edge to edge. Decimals are fine: if one cell spans a quarter of the field, enter 4; if it spans two-thirds of it, enter 1.5.

    For comparison, roughly: a typical rod-shaped bacterium such as Escherichia coli, 1 to 3 µm long; a human red blood cell, 7 to 8 µm across; a human cheek cell, 50 to 60 µm; a Paramecium caudatum, 200 to 300 µm long.

    🧮 Where the Numbers Come From

    Total magnification is objective × ocular. Field of view is the ocular’s field number divided by the objective’s magnification. It does not depend on the ocular’s magnification, because the field number already describes the ocular. If you measured the field instead, the field at any other objective is your measurement scaled by the ratio of the two objectives’ magnifications.

    Resolution is the smallest distance at which two points can still be told apart as two. Following Ernst Abbe, who worked out the theory in the 1870s, it is estimated here as

    d = λ / (NAobjective + NAcondenser)

    where λ is the wavelength of the light and NA is numerical aperture, a measure of how wide a cone of light a lens gathers. With the condenser matched to the objective this becomes the familiar d = λ/2NA. A related and commonly quoted criterion, due to Lord Rayleigh, gives d = 0.61λ/NA, which yields similar values. Both are idealized: real specimens, real lenses, and real alignment generally do somewhat worse.

    Blue filters. Because d is proportional to λ, shorter wavelengths resolve finer detail, and a filter that passes mainly blue light, around 450 to 490 nm, improves resolution by roughly 10 to 20% compared with green; choose “blue filter” above to see the difference. Not every blue filter does this, however. The blue filter found beneath the condenser of many teaching microscopes is often a “daylight” filter, meant mainly to correct the yellowish color of a halogen lamp, and it still passes a broad range of colors, so its effect on resolution is expected to be modest. Blue light also looks dimmer, because the eye is less sensitive to it.

    Numerical aperture is also why oil immersion works. A lens in air cannot have an NA above 1.0, and in practice dry objectives stop near 0.95. Immersion oil has nearly the same refractive index as glass, so light that would otherwise bend away at the glass–air boundary is instead collected by the objective, allowing NAs of 1.25 to 1.4.

    Useful magnification is the range, by a widely used rule of thumb about 500 to 1000 times the objective’s NA, over which magnification enlarges resolved detail enough for your eye to see it comfortably. Below that range the lenses resolve more than you can see; above it, the image gets bigger without getting sharper, what is called empty magnification. With the best oil-immersion objectives, at NA 1.25 to 1.4, the range tops out between about 1250× and 1400×, and that is the reason why light microscopes rarely go past 1000× to 1500×.

    The iris diaphragm trade-off. Closing the iris diaphragm narrows the cone of light coming from the condenser, which lowers NAcondenser. Contrast and depth of field improve, which is why closing it makes transparent cells visible, but d grows, so the finest detail blurs. Try the slider above. The best setting is a compromise, and it differs from specimen to specimen, which is why it is worth adjusting rather than leaving where you found it.

    🔢 Counting Cells: the Petroff-Hausser Chamber

    A counting chamber turns a microscope into a measuring instrument. It is a thick glass slide with a grid etched into a central platform. The platform sits a precise distance below the coverslip, so every square of the grid holds a known volume of liquid. Count the cells over some number of squares, divide by the volume, and you have the number of cells per milliliter.

    The Petroff-Hausser chamber is made for bacteria. Its platform is only 0.02 mm below the coverslip, shallow enough that small cells stay close to a single focal plane. The more familiar hemocytometer uses essentially the same grid at a depth of 0.1 mm, which suits larger cells such as yeast and blood cells. The calculator below handles either.

    🔲 The Grid

    The ruled area of a Petroff-Hausser chamber is a square 1 mm on a side. It is divided into 25 groups of squares, 5 by 5, and each group is divided again into 16 small squares, 4 by 4, for 400 small squares in all.

    Counting unitSideAreaVolume at 0.02 mm deepCells/mL =
    small square0.05 mm1/400 mm²5 × 10−8 mLmean count × 2 × 107
    10 small squares–1/40 mm²5 × 10−7 mLcount × 2 × 106
    group of 160.2 mm1/25 mm²8 × 10−7 mLmean count × 1.25 × 106
    whole ruled area1 mm1 mm²2 × 10−5 mLcount × 5 × 104

    Each result is then multiplied by any dilution made before loading. The conversions rest on 1 mm³ = 1 µL = 10−3 mL.

    The 1 mm ruled area: 25 groups (heavy lines), each of 16 small squares. One group and one small square are shaded.

    🔢 Loading and Counting

    1. Clean the chamber and its coverslip, and place the coverslip over the ruled area.
    2. Mix the sample well, since cells settle. If it is dense, dilute it, and note by how much.
    3. Motile bacteria must be stopped before they can be counted, for example by adding a fixative. A drop of stain, if used, is also a dilution to account for.
    4. Touch a small drop of sample to the edge of the coverslip and let capillary action draw it in. The chamber should fill without overflowing into the grooves around the platform; an overfilled chamber is deeper than it should be.
    5. Give the cells a minute or two to settle, then focus on the grid. Unstained bacteria are seen best with phase-contrast optics (see Types of Microscopy). With brightfield, close the iris diaphragm well down.
    6. Count small squares spread across the grid rather than side by side: for example, 10 small squares scattered over the grid, rather than one full group of 16. Counting more squares, and so more cells, shrinks the counting error, as described below.

    Cells on the lines. A cell touching a boundary could be counted in either of two squares, or in neither. A common convention is to count cells that touch the top or left boundary of a square but not those touching the bottom or right boundary. Whatever convention you use, use it consistently, so that no cell is counted twice. The practice field at the bottom of this tab uses this one.

    Why a few hundred? Cells land in squares at random, so counts vary from square to square even in a perfectly mixed sample. For randomly scattered cells, the uncertainty of a total count N is roughly √N, a relative uncertainty of about 1/√N: 20% for 25 cells, 10% for 100, and 5% for 400.

    Total versus viable. A counting chamber counts every cell it can see, alive or dead. A plate count, by contrast, counts only cells able to form colonies. The two therefore answer different questions, and a chamber count is expected to be at least as high as a plate count of the same sample.

    🧮 Calculator

    1 for undiluted; 10 for 1 part sample plus 9 parts diluent.

    Counting 10 small squares? Either enter the 10 counts, one per square, with “small square” chosen above, or enter their total as a single count, with “several small squares (total)” chosen and set to 10. Entering the counts square by square also lets the calculator check whether they vary more than chance alone would explain.

    🎯 Practice Counting

    Below is one group of 16 small squares with cells scattered over it, some of them on the boundary lines. Count the cells that belong to this group, counting those touching its top or left boundary but not those touching its bottom or right boundary. Then check your count.

    🔍 Types of Microscopy

    Every kind of microscopy has to solve two problems: making small things big enough to see (magnification and resolution) and making them stand out from their surroundings (contrast). Brightfield microscopy, the kind used on the rest of this page, is simple and inexpensive, but it gets its contrast mostly from staining, and staining generally kills cells. The other methods below differ mainly in how they create contrast, what they can resolve, and what they require of the specimen. Resolutions given are rough, typical values.

    💡 Light Microscopy

    Compound brightfield

    Light passes through the specimen, which is seen as darker or colored against a bright background.

    Uses
    Stained tissues and smears, Gram stains, general teaching and clinical work.
    Pluses
    Simple, inexpensive, robust, and quick; resolves down to roughly 0.2 µm with oil immersion.
    Minuses
    Living, unstained cells show little contrast; staining usually requires killing and fixing them.

    Stereo (dissecting)

    Two separate optical paths, one per eye, give a three-dimensional view of the surface of an object, usually lit from above.

    Uses
    Whole insects, flowers, colonies on plates, dissections, and anything needing manipulation under the lens.
    Pluses
    Depth perception, long working distance, and opaque specimens viewed as they are.
    Minuses
    Low magnification, typically up to about 50×; not for individual cells.

    Darkfield

    A special condenser lights the specimen only from oblique angles, so only light scattered by the specimen enters the objective. Specimens glow against black.

    Uses
    Living, unstained microorganisms, including very thin bacteria such as spirochetes that are hard to see otherwise.
    Pluses
    High contrast without staining; inexpensive add-on to a brightfield microscope.
    Minuses
    Little internal detail; dust and debris glow too; needs bright illumination.

    Phase contrast

    Converts the tiny delays that light suffers passing through denser parts of a transparent specimen into differences in brightness. Invented by Frits Zernike, for which he received the 1953 Nobel Prize in Physics.

    Uses
    Living cells in culture, motile microorganisms, bacterial counts in Petroff-Hausser chambers, bacterial spores.
    Pluses
    Internal structure of living, unstained cells; widely available.
    Minuses
    Bright halos around edges; works poorly with thick specimens; requires matched objectives and condenser rings.

    Differential interference contrast (DIC)

    Also called Nomarski microscopy, after Georges Nomarski. Polarized light is split into two slightly offset beams that are recombined after passing through the specimen, giving an image that looks shaded, almost three-dimensional.

    Uses
    Living cells and embryos, especially where fine edges and surfaces matter.
    Pluses
    No halos, thin optical sections, and striking relief-like images.
    Minuses
    Expensive optics; the apparent relief is an optical effect, not true topography; problems with plastic dishes.

    Fluorescence

    Specimens are labeled with molecules that absorb light of one color and emit another, and filters pass only the emitted light. Labels include fluorescent dyes, antibodies carrying dyes, and fluorescent proteins such as GFP.

    Uses
    Locating specific molecules, structures, or cell types; live/dead staining; clinical diagnostics.
    Pluses
    Great specificity and sensitivity; several labels can be imaged in different colors at once.
    Minuses
    Only labeled things are seen; labels fade with exposure (photobleaching); light can harm living cells; out-of-focus glow blurs thick specimens.

    Confocal

    A fluorescence method in which a pinhole blocks light from outside the plane of focus. Scanning the specimen point by point builds up sharp optical sections, which can be stacked into three dimensions. The principle was patented by Marvin Minsky, who filed in 1957.

    Uses
    Thick specimens, tissues, biofilms, and three-dimensional reconstructions.
    Pluses
    Removes out-of-focus blur; optical sectioning without physically cutting the specimen.
    Minuses
    Expensive; slower than widefield imaging; more photobleaching; resolution only modestly better than conventional fluorescence.

    Super-resolution

    A family of fluorescence methods (among them STED, PALM, STORM, and structured illumination) that get around the diffraction limit described on the Calculators tab by controlling when and where labels fluoresce. Eric Betzig, Stefan Hell, and William Moerner shared the 2014 Nobel Prize in Chemistry for this work.

    Uses
    Arrangements of molecules within cells and within bacteria.
    Pluses
    Resolution of roughly 20 to 100 nm, depending on the method, with the specificity of fluorescence.
    Minuses
    Specialized instruments and labels; often slow; demanding sample preparation and analysis.

    ⚡ Electron Microscopy

    Electrons have far shorter wavelengths than visible light, so electron microscopes can resolve far smaller things. The first was built by Ernst Ruska and Max Knoll in the early 1930s; Ruska received a share of the 1986 Nobel Prize in Physics. The price is that specimens must be in a vacuum, which rules out living cells.

    Transmission electron microscopy (TEM)

    Electrons pass through a very thin section or a thin layer of particles. Contrast comes from heavy-metal stains, including negative staining, in which the stain surrounds rather than enters small particles.

    Uses
    Cell ultrastructure, membranes and organelles, and viruses, including bacteriophages, whose shapes are far below the resolution of light microscopes.
    Pluses
    Nanometer-scale resolution of biological specimens.
    Minuses
    Expensive; elaborate preparation that can introduce artifacts; specimens must be dead, dry, and thin; images are two-dimensional slices.

    Scanning electron microscopy (SEM)

    A focused beam of electrons scans across the surface of a specimen, usually coated with a thin layer of metal, and electrons given off from the surface build the image.

    Uses
    Surfaces of cells, tissues, insects, pollen, and biofilms.
    Pluses
    Dramatic three-dimensional-looking images with great depth of field; resolution of a few nanometers or better.
    Minuses
    Surfaces only; drying and coating can distort specimens; black-and-white (color in SEM images is added afterward).

    Cryo-electron microscopy (cryo-EM)

    Specimens are frozen so quickly that water becomes glass rather than ice, and are imaged unstained while frozen. Thousands of images of identical particles can be combined into three-dimensional structures. Jacques Dubochet, Joachim Frank, and Richard Henderson shared the 2017 Nobel Prize in Chemistry for its development.

    Uses
    Structures of proteins, molecular machines, and viruses, including phages; with cryo-electron tomography, the interiors of intact small cells.
    Pluses
    Near-native state without stains; structures often at near-atomic resolution.
    Minuses
    Very expensive instruments; heavy computation; beam damage limits exposure; not for living specimens.

    📍 Scanning Probe Microscopy

    Atomic force microscopy (AFM)

    No lenses or beams at all: an extremely fine tip on a flexible arm is dragged or tapped across a surface, and its deflections map the surface’s height.

    Uses
    Surfaces of cells and molecules, sometimes under liquid; measurements of stiffness and adhesion.
    Pluses
    Nanometer-scale height information, in air or liquid, potentially on living cells.
    Minuses
    Slow; surfaces only; small scan areas; images reflect the shape of the tip as well as the specimen.

    📜 Where It All Began

    Robert Hooke described and named cells, in cork, in his Micrographia of 1665, using a compound microscope. Antonie van Leeuwenhoek, using single-lens microscopes of his own making, soon after became the first to report protists and bacteria. For much of the next two centuries, however, compound microscopes were limited more by the imperfections of their lenses than by the physics of light, and it was not until Abbe’s work in the 1870s that the ultimate limit of light microscopy, described on the Calculators tab, was understood.

    🧮 Phage Biology and Phage Therapy Calculators